Page 45
Control Commands
optoNCDT ILR 1191
10.3.7 SO – Set Offset
Format: SO [Enter]
The command performs one single distance measurement, then sets it as –OF (offset). Function SO can only
be carried out. It does not represent a parameter in the actual sense of the word.
Use the command SO for example for doing measurements.
Output value [m] = Measured value - OF
Use the command OF0 for finishing the relative measurement.
4 mA
10 mA
15 mA
20 mA
QA0 210
after Offset
QA60 210
after Offset
0 m
30 m
150 m
69 m
Distance
Signal
Fig. 18 Analog output with offset displacement at 69 m