Page 53
Setting Sensor Parameters
optoNCDT 1750
7.3
Inputs
Change to the
Input
menu in the
Settings
tab
.
Laser power
Full / Reduced / Off
The laser light source is active only, if pin 9 is
connected to GND.
Synchronization
Slave /
Slave alternating
Termination
On / Off
If several sensors measure the same target
synchronously, the sensors may be synchro-
nized with each other. The synchronization
output of the first sensor (master) controls
the sensors connected to the synchronization
inputs (slaves).
Slave MFI
Master / Master alternating
inactive
Level Multi-func-
tion input
TTL / HTL
Defines the input level of both switching in-
puts
Laser on/off
and
Multi-function
.
TTL: Low ≤ 0,8 V; High ≥ 2 V
HTL: Low ≤ 3 V; High ≥ 8 V
Grey shaded
fields require a
selection.
Value
Dark-bordered
fields require
you to specifie
a value.
Содержание optoNCDT 1750