Page 70
Setting Sensor Parameters
optoNCDT 1750
7.6.2
Digital Output, RS422
7.6.2.1
Values, Ranges
The digital measurement values are issued as unsigned digital values (raw values). 16 or 18 bits can be transferred per value. Below
you will find a compilation of the output values and the conversion of the digital value.
Value
Length Variables
Value range
Formula
Distance
18 Bit x = digital value
[0; 230604]
d
[mm]
=
* MR
[mm]
x - 98232
65536
MR = Measuring range [mm] {2/10/20/50/100/200}
d = Distance [mm]
without mastering [-0.01MR;
1.01MR]
with mastering [-2MR; 2MR]
Exposure time 16 Bit x = digital value
[1334; 33333]
ET
[µs]
=
x
1
10
ET = Exposure time [µs]
[133.4; 3333.3]
Intensity
16 Bit x = digital value
[0; 1023]
I
[%]
=
x
100
1023
I
= Intensity [%]
[0; 100]
Sensor state
18 Bit x = digital value
[0; 242143]
Bit 0 (LSB): peak starts before ROI
Bit coding
[0; 1]
Bit 1: peak ends after ROI
Bit 2: no peak found
SMR =Start of measuring range
Bit 5: Distance before SMR (extended)
EMR = End of measuring range
Bit 6: Distance after EMR (extended)
Bit 15: Measurement value is triggered
Bit 16, 17: Status LED;
- 00 – off
10 – red
- 01 – green
11 – yellow