Appendix | ASCII Communication with Sensor
optoNCDT 1220
A 3.3.8 Measurement
A 3.3.8.1 MEASRATE, Measuring Rate
MEASRATE 0.25|0.5|1
Choice of measuring rate in kHz.
A 3.3.8.2 LASERPOW, Laser Power
LASERPOW FULL|OFF
- FULL: laser power is set to 100 %
- OFF: laser is switched off
A 3.3.8.3 MASTERMV, Mastering / Zeroing
MASTERMV NONE|MASTER <MV>
- NONE: completes mastering
- MASTER: sets the current measurement value as a master value
- MV: master value in millimeters; MV = (0 ... 2) * measuring range, i.e. master value must be within measur-
ing range
In case of master value is 0, the mastering has the same functionality as the zeroing. The parameter MV
always operates independently from an input as zeroing during mastering the analog output.
The master command awaits the next measurement value, a maximum of 2 seconds, and masters it. If no
measurement value is received within this time, for example, by external triggering, the command returns with
the error “E220 Timeout“.
The master value is processed with six decimal places.
Note that the output value is limited to 18 bits.