Page 16
Functional Principle, Technical Data
optoNCDT 1220
3.3
Technical Data
Available for OEM series
Model
ILD1220-10
ILD1220-25
ILD1220-50 ILD1220-100 ILD1220-200
ILD1220-500
Measuring range
10 mm
25 mm
50 mm
100 mm
200 mm
500 mm
Start of measuring range
20 mm
25 mm
35 mm
50 mm
60 mm
100 mm
Mid of measuring range
25 mm
37.5 mm
60 mm
100 mm
160 mm
350 mm
End of measuring range
30 mm
50 mm
85 mm
150 mm
260 mm
600 mm
Measuring rate
1
3 adjustable stages: 1 kHz / 0.5 kHz / 0.25 kHz
Linearity
< ±10 µm
< ±25 µm
< ±50 µm
< ±100 µm
< ±200 µm
< ±750 ... 1500 µm
< ±0.10 % FSO
< ±0.15% ... 0.30 % FSO
Repeatability
2
1 µm
2.5 µm
5 µm
10 µm
20 µm
50 µm
Temperature stability
±0.015 % FSO / K
±0.01 % FSO / K
Light spot diameter
(±10%), [µm]
SMR
90 x 120
100 x 140
90 x 120
750 x 1100
MMR
45 x 40
120 x 130
230 x 240
EMR
140 x 160
390 x 500
630 x 820
smallest
Ø
45 x 40
with 24 mm
55 x 50
with 31 mm
70 x 65
with 42 mm
-
Light source
Semiconductor laser < 1 mW, 670 nm (red)
Laser safety class
Class 2 in accordance with IEC 60825-1: 2014
Permissible ambient light
3
20,000 lx
7,500 lx
Supply voltage
11…30 VDC
Power consumption
< 2 W (24 V)
Analog output
4 ... 20 mA (12 bit, freely scalable within the measuring range)
4