Page 103
Anhang | API Communication with Controller (Version 0.7.1)
colorSENSOR CFO
A 3.5.3.7 /api/firmware/images/UUID/apply
POST
Application of firmware image (permanent transfer to internal data storage medium) followed by sensor
restart
Error codes:
LPLC.format.malformed.upload
LPLC.internal_error
A 3.5.3.8 /api/firmware/images/UUID/upload
The firmware image is uploaded in data blocks (see
max_chunk_size
). The HTTP header
Content Range
is used by the client to notify the server of the range to be transferred with the current POST request. The
transfer is accomplished without gaps (i.e. the start address for the subsequent block corresponds to the end
address for the previous block incremented by 1).
It is permissible to upload additional blocks as long as the firmware image status is
incomplete
. All other
statuses indicate successful or erroneous conclusion of the firmware upload. In the event of a permanent
error status for the upload operation (e.g. erroneous check sum), each request is acknowledged with the
Status Code 400 (
Bad Request
).
Error codes:
LPLC.header.content_range.missing
LPLC.header.content_range.invalid
LPLC.header.content_range.conflicts
LPLC.upload.missing_chunk
LPLC.upload.payload_too_big
LPLC.format.malformed.upload
Содержание colorSENSOR CFO Series
Страница 1: ...CFO100 CFO200 Instruction Manual colorSENSOR CFO...
Страница 6: ...colorSENSOR CFO...
Страница 69: ...Page 69 Operation colorSENSOR CFO Fig 27 Screen shot color table...
Страница 81: ...Page 81 Operation colorSENSOR CFO 3 2 1 5 6 7 8 4 Fig 34 Measuring data display...
Страница 121: ......
Страница 122: ......
Страница 123: ......