Page 47
Commands
colorCONTROL MFA-5
6.2.2.2 Manual Check
capturexyz b
Measures and saves the color and intensity of the individual LEDs with the specified test time.
Command
Description
Received
Example
Note
capturexyz b x = test time preselection
1 = 600 ms
2 = 200 ms
3 = 120 ms
4 = 60 ms
5 = 20 ms
6 = 10 ms
7 = 2 ms
8 = can be programmed by the user
9 = current setting will be applied
0 = sensor off
y = 0 color chip Low (3x3) preselec-
tion
y = 1 color chip High (9x9) preselec-
tion
z = preselection channel 1 ... 5 or 1
... n
b = preselection board 1 ... n, only
if z = 1 ... 5
OK
capture 215 3
OK
capture 31
OK
capture 3117
capture 312 4
OK
x = test time 200 ms
y = color chip High
(9x9)
Z = channel 5
b = board 3
x = test time 120 ms
y = color chip High
(9x9)
for all channels and
colorCONTROL
MFA-5
x = test time 120 ms
y = color chip High
(9x9)
Channel 17 or chan-
nel 2 on board 4
This command enables optimum setting for every LED to be tested.
Dark LEDs are checked with a longer test time (e.g. 600 or 200 ms) and the Sensor High setting is used (in
doing so, all 9x9 segments of the color chip are used).
For very bright LEDs, the Sensor Low setting is selected (only 3x3 segments of the color chip are used) and
the test time is reduced accordingly (e.g. 10 or 2 ms).
Board = colorCONTROL MFA-5
Channel = checkpoint
Содержание colorCONTROL MFA-5
Страница 1: ...Instruction Manual colorCONTROL MFA 5...
Страница 26: ...Page 26 Operation colorCONTROL MFA 5 Test single color chip Fig 18 Settings for test sensitivity...
Страница 35: ...Page 35 Operation colorCONTROL MFA 5 Fig 23 Sensor values and intensity section...
Страница 38: ...Page 38 Operation colorCONTROL MFA 5 Fig 26 Comparison test...
Страница 78: ......
Страница 79: ......