Page 17
Installation
colorCONTROL MFA-5
4.3.1
Individual colorCONTROL MFA-5 Sensor
The colorCONTROL MFA-5 can be operated both via an RS232 as well as via a USB port. In USB operation,
the power for a sensor is supplied via the USB interface.
In RS232 operation, an external power supply of 10 - 30 VDC, approximately 80 mA (max. 320 mA) must be
used.
colorCONTROL
Daisy-Chain
USB
RS232
Power
Fig. 10 colorCONTROL MFA-5 housing with connections
USB connection
1
2
3
4
Pin
Description
1
GND
2
VBUS (+5 V)
3
Data
4
Data
Fig. 11 Socket M9 Binder 712
Using USB mode may lead in especially noisy electromagnetic environment to functional interference in indi-
vidual cases. Here, a system reboot may be required.
NOTICE