If E7 occurs when the output is first being
adjusted, it may mean the electrodes or cables are
not making a good circuit. Check cable and
electrode connections and make sure electrodes
are making good contact with the patient.
In the continuous treatment modes the output
voltage is reduced while the unit monitors the
impedance of the patient connection. If the unit is
in amplitude modulated modes, such as recip or
surge, this patient connection error causes the unit
to go into the HOLD mode.
All patient connection errors should be
investigated to determine their cause.
4.
“E80_” displayed in Time
window.
An output overcurrent has been detected. Current
exceeded 70 mA RMS for interferential, 55 mA for
premodulated, and medium frequency or 105 mA
peak for biphasic.
Reposition electrodes farther apart. Remove any
moisture or gel from between the electrodes and
try again. If error persists even without a patient
connection or load, unit requires servicing.
5.
“E90_” displayed in Time
window.
Output error for electrical stimulation has been
detected. Remove electrode cables from unit and
restart. Replace electrode cables onto unit.
Reprogram treatment and try starting treatment
session again.
If powering unit OFF and restarting does not
remove error, the unit requires servicing.
6.
“F1__” displayed in Time
window.
There has been a communication error between
the microprocessors. If powering unit OFF and
restarting does not remove error, unit requires
servicing.
7.
“F2__” displayed in Time
window.
There has been a self-test error— If powering unit
OFF and restarting with all the electrode cables
removed does not remove error, the unit requires
servicing.
8.
“F3 _” displayed in Time
window.
There has been a power supply error— If
powering unit OFF and restarting does not
remove error, the unit requires servicing.
If problem is not addressed above, or if additional troubleshooting guidance is desired,
call (800) 854-9305 or email service directly at
.
The distributor who sold the Sys*Stim 294 should be able to assist you with a loaner unit
during warranty service.
Содержание Sys*Stim 294
Страница 2: ...Mettler Electronics Corp Rev F_09 13 17 2 ...
Страница 18: ...Mettler Electronics Corp Rev F_09 13 17 18 ...
Страница 28: ...Mettler Electronics Corp Rev F_09 13 17 28 ...
Страница 32: ...Mettler Electronics Corp Rev F_09 13 17 32 ...
Страница 42: ...Mettler Electronics Corp Rev F_09 13 17 42 ...