![Mettler Electronics Sys*Stim 294 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/mettler-electronics/sys-stim-294/sys-stim-294_instruction-manual_1780253019.webp)
Sys*Stim 294 Instruction Manual—
Rev.F_09/13/17
19
Section 4—Operating
Instructions
1 5. 0 0
min s
s
s
Hz
1
3
2
4
1 - 15 Hz
80 - 150 Hz
1 - 150 Hz
Hz - Hz
1
2
Hz
Hz
s
Hz
Hz
+
–
+
–
1
2
3
4
5
6
7
8
9
0
1
2
3
4
00 0
00 0
V
mA
A
mA
Stim Channels
Timer Functions
Start
Hold /
Enter
Stop All Output
Stim Setup
Sweep Frequency
Interferential
Amplitude
Modulation
Premodulated
Medium
Frequency
Biphasic
High Volt
MicroCurrent
Phase
Duration
Frequency
Polarity
Frequency
Polarity
Frequency
Continuous
Surge
Reciprocation
Stim
Output
WARNING: Hazardous electrical output. This
equipment is for use by qualified personnel.
Use electrode cable set ME 2260.
!
SYS STIM 294
*
®
Figure 4.1—Front membrane panel and LED indicators
4.1 A Note About Electrodes
To ensure safe operation of the Sys*Stim 294, follow the recommendations listed below:
1.
We strongly encourage careful maintenance of the electrode system. This includes the
lead wires as well as the pads themselves. Worn cables and/or poor pads (or the
wrong sized pads) can have a significant impact upon treatment results.
2.
For best performance, use Mettler Electronics’ brand electrodes.
3.
Do not exceed the number of recommended uses listed on the instructions for EZ
Trodes or other reusable self–adhesive electrodes.
4.
Make sure that the entire surface of the electrode is contacting the patient.
5.
Do not use moist hot packs to secure electrodes.
6.
To avoid skin irritation due to high current density, do not use electrodes smaller in
surface area than the 2" diameter EZ Trode® self-adhesive electrode (ME 2221).
7.
Do not use conductive carbon electrodes with this product.
8.
Whenever clinically possible, utilize the largest possible pads to reduce local increases
in current density. In situations where small pads are required, use the lowest
stimulation intensity necessary to achieve the desired clinical results.
Содержание Sys*Stim 294
Страница 2: ...Mettler Electronics Corp Rev F_09 13 17 2 ...
Страница 18: ...Mettler Electronics Corp Rev F_09 13 17 18 ...
Страница 28: ...Mettler Electronics Corp Rev F_09 13 17 28 ...
Страница 32: ...Mettler Electronics Corp Rev F_09 13 17 32 ...
Страница 42: ...Mettler Electronics Corp Rev F_09 13 17 42 ...