www.metron-biomed.com
15
Remote Control
This is the screen that is shown when the daeg is in
remote control mode. All currently active parameters are
displayed on the screen. Pressing (
F1
) will cancel the
remote control mode and take you back to the local
Main
Simulation Screen
.
Probe Analyzer
This screen will let you select which test you wish to
perform on the finger probe when you have the probe
tester adaptor box attached.
Diode Test
Pressing (
DIODE
) runs the test.
This test gives a descriptive message as to the condition
of the probe’s LEDs and sensing elements. It tests the
probe for electrical shorts and open circuits.
------ STATUS ------
Preset:*Manual
SpO2 : 96 %
Rate : 60 BPM
P.ampl: 100 %
LOCAL
CONTR.
*
F3
F2
F1
-- PROBE ANALYZER –-
Choose a test to
perform on the
SpO2 probe.
DIODE CONTI SENSI
F3
F2
F1
-- PROBE ANALYZER –-
Probe OK!
DIODE CONTI SENSI
F3
F2
F1