Function Selection Menu
This screen will let you select which testing function of
the daeg that you wish to perform. Using the
Up
and
Down
arrows will let you move the cursor (
->
) next to the
function that you wish to use. Once the cursor is placed
press the (
F2
) button or the (
yes/enter
) button to enter
that function. Pressing the (
F3
) or (
esc
) buttons will take
you back to the
Main Simulation Screen
.
Alarm Limits
This test is designed to test your alarm system on your
monitor. The (
F2
) button will move the cursor, (
*
), to the
next parameter. Using the Up and Down arrows will let
you change the value of the selected parameter. You will
then need to press the (
yes/enter
) button to accept the
change and start the timer. Press the (
F1
) button to stop
the clock when you hear the alarm from your monitor.
The (
F3
) button will generate a line of printout via the
parallel port. Pressing the (
esc
) button will take you back
to the
Main Simulation Screen
.
-----FUNCTIONS------
Alarm limits
Print header
Print Status
Remote Contr
ENTER EXIT
F3
F2
F1
-----FUNCTIONS------
Print Status
Remote Contr
Probe Analyz
Sim. Setup
ENTER EXIT
F3
F2
F1
AL_LIMIT t= 0.0 sec
Pulse :*On
SpO2 : 96 %
Rate : 60 BPM
P.ampl: 100 %
STOPt
*
F3
F2
F1
Содержание daeg SPO2
Страница 1: ...daeg SPO2 Simulator User Manual Version 1 15_2 Updated 22 01 2007...
Страница 6: ...www metron biomed com 6 This page was intentionally left blank...
Страница 8: ...www metron biomed com 8 This page was intentionally left blank...
Страница 10: ...www metron biomed com 10 This page was intentionally left blank...
Страница 12: ...www metron biomed com 12 This page was intentionally left blank...
Страница 18: ...www metron biomed com 18 This page was intentionally left blank...
Страница 20: ...www metron biomed com 20 This page was intentionally left blank...