Simulation Setup
This function lets you manually alter the R-Curves within
the daeg unit.
Be cautious when using this mode as you are altering the
simulator output. If you make any changes that you do
not wish to keep you must turn off the daeg for 5 seconds
then turn it back on.
When in this mode the controls will be as follows:
Up
and
Down
arrows control the cursor position.
(
yes/enter
) increases the indicated value.
(
no/clear
) decreases the selected value.
(
F1
) decreases the selected SPO2 value.
(
F2
) increases the selected SPO2 value.
(
F3
) Saves the changes and takes you back to the
Main
Simulation Screen
.
The saving procedure may take up to 5 minutes. Do not
turn off the power during this time.
SIM SETUP:
To change values,
use 'Yes' –button
for up, and
'No' –button for
down!
F3
F2
F1
Make :Nellcor
RedDC:2048
IRDC :1024
Lcode: 101
Rval :0.48@SpO2:100
SAVE &
SpO2- SpO2+ QUIT
F3
F2
F1
Содержание daeg SPO2
Страница 1: ...daeg SPO2 Simulator User Manual Version 1 15_2 Updated 22 01 2007...
Страница 6: ...www metron biomed com 6 This page was intentionally left blank...
Страница 8: ...www metron biomed com 8 This page was intentionally left blank...
Страница 10: ...www metron biomed com 10 This page was intentionally left blank...
Страница 12: ...www metron biomed com 12 This page was intentionally left blank...
Страница 18: ...www metron biomed com 18 This page was intentionally left blank...
Страница 20: ...www metron biomed com 20 This page was intentionally left blank...