4 Maintenance
Spectrosense, Instructions for Use
31
The output signal
is too low.
•
Check if the mirror is tightly screwed onto the
Spectrosense.
•
Check if there are any foreign materials or air
bubbles in the optical path.
•
Check the mirror surface. If it is scratched, replace
the mirror; order number 6.1250.010.
•
Maybe a voltage adjustment of the Spectrosense
has to be done, see
The output signal
is too high
(> 1200 mV).
•
Maybe a voltage adjustment of the Spectrosense
has to be done, see