6 Troubleshooting
884 Professional VA
■■■■■■■■
79
Problem
Cause
Remedy
The measuring solution has
been contaminated as a
result of diffusion.
No infeed tubing other than the four-way
micro dosing tip (6.1824.000) may be
immersed in the measuring solution.
The electrode tip of
the working elec-
trode and/or the
auxiliary electrode
shows copper
deposits.
An incorrect potential has
been applied because the
reference electrode is
defective.
■
Check the reference electrode; perform
maintenance procedure in accordance with
the electrode leaflet, if necessary.
■
Dip the electrode tip of the working elec-
trode and/or the auxiliary electrode into
concentrated nitric acid for 1 to 2 seconds
and then thoroughly rinse with distilled
water.
An incorrect potential was
applied because the refer-
ence electrode is not con-
nected.
■
Check the electrode connectors.
■
Dip the electrode tip of the working elec-
trode and/or the auxiliary electrode into
concentrated nitric acid for 1 to 2 seconds
and then thoroughly rinse with distilled
water.
An incorrect potential was
applied because the meas-
uring vessel contains too
little measuring solution.
■
All three electrodes must be immersed in
the measuring solution.
■
Dip the electrode tip of the working elec-
trode and/or the auxiliary electrode into
concentrated nitric acid for 1 to 2 seconds
and then thoroughly rinse with distilled
water.
The measurement
curves are noisy.
Contact problem at the
driving axle (6.1204.510 or
6.1204.520).
■
Remove the abrasion residue from the driv-
ing axle.
■
Replace the driving axle.
The measuring sig-
nal for Cu VMS fluc-
tuates.
The ambient and/or solu-
tion temperature is not
constant.
Keep the ambient and solution temperature
stable during measurements (±4 °C).
The reference potential
drifts off.
Perform the maintenance procedure for the
reference electrode in accordance with the
electrode leaflet. However, for CVS analyses,
replace the reference electrolyte every other
day and the bridge electrolyte every day. After
replacing the reference electrolyte, wait at
least 20 minutes (preferably 1 - 2 h) until the
potential has reached equilibrium.