5.1 Replacing the measuring head
■■■■■■■■■■■■■■■■■■■■■■
66
■■■■■■■■
884 Professional VA
5
CAUTION
Do not apply pressure to the drive disk on the connector plate of
the measuring head arm, as this could damage the stirrer motor.
Figure 43
Do not touch the drive disk
Use one hand to hold the rear of the measuring head arm and use
the other hand to insert the new measuring head into the connector
plate of the measuring head arm.
The measuring head must snap into place with an audible click.
6
WARNING
If the measuring head arm is folded down without due care, this
may result in injuries to the hands.
Make sure that your fingers do not get caught between the meas-
uring head arm and the instrument housing.
Lower the measuring head arm again.