3.2 Equipping the MME measuring head
■■■■■■■■■■■■■■■■■■■■■■
32
■■■■■■■■
884 Professional VA
NOTE
Please also observe the notes contained in the electrode leaflets sup-
plied with the electrodes. In addition, you can learn how to best handle
the electrodes from the multimedia guide (A.717.0002).
For detailed information about handling the Multi-Mode Electrode pro
(MME pro), observe the
Multi-Mode Electrode pro
document
(8.110.8018XX).
3.2.2.1
Working electrode (WE)
Preparing and inserting the working electrode
Proceed as follows:
1 Preparing the working electrode
Prepare the working electrode in accordance with the instructions in
the
Multi-Mode Electrode pro
document (8.110.8018XX). In prepa-
ration, carry out the following tasks:
■
Mount the capillary
■
Insert the needle
■
Topping up the mercury
2 Inserting the working electrode into the measuring head
insert
■
Place an empty measuring vessel in the holder
■
Carefully insert the working electrode into the opening
of
the measuring head insert. The bottom of the capillary must not
touch the measuring head while being inserted.