![Metrohm 799 GPT Titrino Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/metrohm/799-gpt-titrino/799-gpt-titrino_short-instructions-for-use_1779570024.webp)
2.9
Mode TIP, parameters
799 GPT Titrino: Titration, dosing and measuring modes
22
2.9 Mode TIP, parameters
With TIP, methods and other functions can be linked to a ti-
tration sequence. Press the key <DEF> to define the titra-
tion sequence.
Display Meaning
Input
range
>sequence
method:
pause
L4 output:
L6 output:
info
prep
stirrer:
Method from the internal method memory
or from the card.
Waiting time. Can be aborted with
<QUIT>.
Set a signal when on I/O line L4.
Set a signal when on I/O line L6.
The sequence will be held and an info
appears in the display.
Preparation of the dosing unit.
Switching the stirrer ON/OFF.
method name
0
...999 999 s, inf.
active, inactive, pulse,
OFF
active, inactive, pulse,
OFF
up to 16 characters
internal D0
, external D1/D2
ON
, OFF
>statistics
Statistics calculation see at DET, page 14.
>preselections
req.ident:
req.smpl size:
limit smpl size:
meas.mode:
meas.input:
temperature
Request of identifications after start.
Request of sample size after start.
Limit control of sample size.
Selection of the measured quantity for
<MEAS/HOLD>.
Measuring input for pH and U.
Temperature for the compensation of the
pH values.
id1, id1 & 2, all,
OFF
value, unit, all,
OFF
ON,
OFF
pH, U, I(pol), U(pol), T,
OFF
1
, 2, diff.
-170.0...500.0 °C