2.8
Mode MEAS and CAL, parameters
799 GPT Titrino: Titration, dosing and measuring modes
21
2.8 Mode MEAS and CAL, parameters
MEAS
Display Meaning
Input
range
>measuring parameters
signal drift
equilibr.time
meas.input:
temperature
time interval
Drift for measured value acquisition.
Waiting time for meas. value acquisition.
Measuring input for pH and U.
Measuring temperature.
Time interval for measured value
acquisition.
pH, U: 0.5..999 mV/min,
OFF
T: 0.5...999 °C/min,
OFF
0...9999 s,
OFF
1
, 2, diff.
-170.0...
25.0
...500.0 °C
1...
2
...999 999s
>statistics
Statistics calculation see at DET, page 14.
>preselections
req.ident:
req.smpl size:
Limit smpl size:
activate pulse:
Request of identifications after start.
Request of sample size after start.
Limit control of sample size.
Pulse output on I/O line L6.
id1, id1 & 2, all,
OFF
value, unit, all,
OFF
ON,
OFF
ON,
OFF
CAL
Display Meaning
Input
range
>cal. parameters
meas.input:
cal.temp.
buffer 1 pH
signal drift
equilibr.time
electr.id
sample changer cal:
activate pulse:
Measuring input.
Calibration temperature.
pH-value for buffer 1. up to 9 buffer.
Drift for measured value acquisition.
Waiting time for meas. value acquisition.
Electrode identification.
Calibration with Sample Changer.
Pulse output on I/O line L6.
1
, 2, diff.
-20.0...
25.0
...120.0 °C
0...
±
20.00
0.5..
2
...999 mV/min, OFF
0...
110
...999 s, OFF
up to 8 ASCII characters
ON,
OFF
all, first,
OFF
>statistics
Statistics calculation see at DET, page 14.