2.4
Mode KFT, parameters
799 GPT Titrino: Titration, dosing and measuring modes
16
2.4 Mode KFT, parameters
Display Meaning
Input
range
>control parameters
EP at ...
dynamics
max.rate
min.volume incr.
stop crit:
stop drift
t(delay)
stop time
Controlling parameters for KFT.
Endpoint for Ipol.
Endpoint for Upol
Distance from EP where constant dosing
stops and controlling begins.
Maximum dosing rate.
Minimum volume increment
Type of stop criteria.
Titration stops if stop drift is reached.
Stop if there is no dosing during t(delay).
If t(delay) is ”INF”: stop time.
U: -2000 ...
250
...2000 mV
I: -200.0...
25.0
...200.0 µA
U: 1...
100
...2000 mV
I: 0.1...
10
...200.0 µA
0.01...150 mL/min,
max
.
0.1...9.9 µL,
min
drift
, time
1...
20
...999 µL/min
0...
10
...999 s, INF
0...999 999 s,
OFF
>titration parameters
titr.direction:
pause 1
start V:
start V
factor
dos.rate
pause 2
extr.time
dos.element
I(pol)
U(pol)
electrode test:
temperature
time interval
General titration parameters.
+: Titration to higher voltage or current.
auto: Direction is set automatically.
Waiting time
before
start volume.
Type of start volume: absolute or relative.
Volume for
absolute
start volume.
Factor for
calculation of
relative
start
volume: factor * smpl size.
Dosing rate for start volume.
Waiting time
after
start volume.
Extraction time.
Selection of the dosing unit.
Polarization current for Ipol or -
voltage for Upol in 10 mV-intervals.
Performing of electrode test
Titration temperature.
Time interval for measured value
acquisition.
+,
-
, auto
0
...999 999 s
abs., rel.,
OFF
0
...999.99 mL
0
...±999 999
0.01...150 mL/min,
max.
0
...999 999 s
0
...999 999 s
internal D0
, external D1/D2
-127...
50
...127 µA or.
-1270...
400
...1270 mV
ON,
OFF
-170.0...
25.0
...500.0 °C
1...
2
...999 999 s
>stop conditions
stop V:
stop V
factor
filling rate
Type of stop volume: absolute or relative.
Volume for
absolute
stop volume.
Factor for
calculation of
relative
stop
volume: factor * smpl size.
Filling rate after the titration.
abs.
, rel.,
OFF
0...
99.99
...9999.99
mL
0...±
999 999
0.01...150 mL/min,
max.
>statistics
Statistics calculation see at DET, page 14.
>preselections
conditioning:
display drift:
drift corr:
drift value
req.ident:
req.smpl size:
limit smpl size:
oven:
activate pulse:
Automatic conditioning of titration vessel.
Display of drift during conditioning.
Type of drift correction.
Value for manual drift correction.
Request of identifications after start.
Request of sample size after start.
Limit control of sample size.
RS-COM of Titrino where Oven is
connected.
Pulse output on I/O line L6.
ON,
OFF
ON
, OFF
auto, man.,
OFF
0.0
...99.9 mL/min
id1, id1 & 2, all,
OFF
value, unit, all,
OFF
ON,
OFF
COM1, COM2,
OFF
first, all, cond.,
OFF