2.1
Survey of the titration modes
799 GPT Titrino: Titration, dosing and measuring modes
13
If you are looking for more information to the different quanti-
ties, you will find the display text as a keyword in the index of
the Instructions for Use.
SET, KFT
Endpoint Titration
STAT
Holding a measured value
DOC
Controlled dosing
Titration to preset end-point.
The reagent addition is controlled
such that a preset measured
value is adhered to as accurately
as possible.
The dosing function is controlled
such that a nominal measured
value is adhered to as accurately
as possible. The nominal meas-
ured value follows a linear ramp.
Volume that has been dispensed
up to the endpoint (EPX in mL).
Evaluation of
•
rates dV/dt in the set time
windows and the mean rate
over the entire measuring
point list.
•
volumes associated with a
preset time (fix V).
•
times associated with a
preset fraction of the total
volume (fix time).
The total volume (C41) and the
time needed for dosing (C42)
can be used for calculation.
•
For rapid, quantitative deter-
minations in analytical chem-
istry. Requirement: EP of the
titration reaction is known
and does not change during
a determination series.
•
If an excess of titrant must be
avoided.
•
KFT:
Determinations of water
contents.
For
•
the controlled dosing to a
preset endpoint in the syn-
thesis laboratory.
•
activity determinations, e.g. of
enzymes.
•
kinetic investigations (first
order kinetics).
Sweeping a measured value
gradient for reaction formulations
in the syntheses laboratory.
V/mL
U/mV
continuously
Control
range
EP
t/s
V/ml
rate
t/s
U/mV
V/ml
U
set