![Metrohm 757 VA Computrace Скачать руководство пользователя страница 54](http://html.mh-extra.com/html/metrohm/757-va-computrace/757-va-computrace_hardware-manual_1779655054.webp)
3 Installation
757 VA Computrace – Hardware
50
3.9.4
Operation of the 813 Compact Autosampler
After installation of the instruments according to
sections
3.9.1
...
3.9.3
sample series
using the 813 Compact Autosampler can be started. Proceed always in the follow-
ing sequence:
1 Switch on instruments
•
Switch on Dosimats and 757 VA Computrace Stand.
•
Switch on 813 Compact Autosampler and 731 Relay Box.
•
Switch on PC.
•
Start 757 VA Computrace software (see
Software Manual
,
section 2.2
).
2 Load and modify method
•
Click on
or
MAIN WINDOW
/ Mode / Determination
.
•
Click on
or
MAIN WINDOW
/ Window / Working method specification
to
open the
WORKING METHOD SPECIFICATIONS
window.
•
Load the desired method in the
WORKING METHOD SPECIFICATIONS
window.
•
If desired, modify the loaded method (details see
Software Manual
).
3 Load and modify sample table
•
Click on
or
MAIN WINDOW
/ Window / Sample table
to open the
SAMPLE
TABLE
window.
•
Load the desired sample table or edit the current sample table (details
see
Software Manual
).
4 Place samples on 813 Compact Autosampler
•
Transfer the desired sample amount (5...10 mL) into the sample vessels
and place them at the odd positions on the sample rack of the 813 Com-
pact Autosampler.
•
For each sample vessel, place a vessel filled with rinsing solution at the
following even position (volume rinsing solution = volume sample solu-
tion).
5 Start determination
•
Click on
or
MAIN WINDOW
/ Window / Monitor
to open the
MONITOR
window.
•
Start the measurement by clicking the
icon in the
MAIN WINDOW
or
the
button in the
MONITORING
window.
•
Press the "START" button on the keypad of the 813 Compact Autosam-
pler.
•
Follow the instructions in the appearing message windows.