3 Installation
757 VA Computrace – Hardware
36
3.5 Reference
electrode
3.5.1 Construction
The complete reference electrode (RE)
22
comprises two parts:
•
6.0728.0X0 Ag/AgCl reference system (
101
)
with ceramic diaphragm type D, diameter = 1 mm
6.0728.020
Reference system: Ag/AgCl/
c
(KCl) = 3 mol/L;
supplied in a holder filled with
c
(KCl) = 3 mol/L
as
standard
6.0728.010 Reference system: Ag/AgCl
supplied dry (option)
•
6.1245.010 Electrolyte vessel (
102
)
with ceramic diaphragm type D, diameter=3 mm; holds
a second electrolyte solution (bridge electrolyte) and
thus forms with the 6.0728.020 Reference system a
complete reference electrode in the so-called double
junction construction.
The construction of the reference electrode and the designations of the individual
parts are shown in
Fig. 13
.
22
102
109
110
111
101
103
104
105
106
107
108
Fig. 13: Construction of the reference electrode
22
Reference electrode
101
Reference system (6.0728.0X0)
102
Electrolyte vessel (6.1245.010)
103
Electrical connection for cable
"AE"
104
Vent opening
105
Ag/AgCl filling
106
Electrolyte compartment with
internal electrolyte
107
Diaphragm support made of
PCTFE
108
Diaphragm
109
Vent opening
110
Electrolyte compartment with
bridge electrolyte
111
Diaphragm