
3.9 Connection of the 813 Compact Autosampler
757 VA Computrace – Hardware
49
2 Set automation parameters
•
Click on
MAIN WINDOW
/ Settings / General settings
and select the
Automa-
tion
tab.
•
Enable the
Use Autosampler
option, modify the automation parameters as
desired and close the
GENERAL SETTINGS
window.
•
Close the "757 VA Computrace 2.0" program and restart it.
3 Test automation parameters
•
Fill two sample vessels with water and place them one after the other on
the sample rack of the 813 Compact Autosampler.
•
Click on
MAIN WINDOW
/ Settings / General settings
and select the
Automa-
tion
tab.
•
Click on
, check the automation parameters and modifiy
them if need be. The default settings can be normally used for samples
with 10 mL volume: