Assessing a metal electrode
■■■■■■■■■■■■■■■■■■■■■■
10
■■■■■■■■
Dosing rate
Max.
Stop volume
3 mL
Stop EP
9
Filling rate
Max.
EP criterion
5
EP recognition
All
4
Compare the measurement result with the following specifications:
■
Consumption (EP) [mL]:
1.95–2.05
■
Potential jump [mV]:
deltaU
90–110%
> 70 mV
■
Titration time [s]:
approx. 150
NOTICE
If the measurement result does not correspond to the specifica-
tions, clean the electrode and run the test again. If titration
times are excessive, the stirring rate and the arrangement of the
stirrer, pipetting tip and electrode should be checked.