Storing the dAg ring electrode
■■■■■■■■■■■■■■■■■■■■■■
4
■■■■■■■■
■
Hold the electrode and storage vessel firmly in your hand so that
the electrode cannot slip away.
■
Position the tool between the storage vessel and SGJ sleeve.
■
Carefully
push the tool to the side to release the electrode.
Do not tip the tool forwards!
NOTICE
Avoid applying excess pressure to the tool. Otherwise, the elec-
trode could be released too abruptly.
3 Checking the electrode for proper function
■
Preparing the electrode:
Preparing the dAg ring electrode (see chapter 4.1, page 6)
■
Checking the electrode:
Assessing a metal electrode (see chapter 5.2, page 9)
NOTICE
Defective electrodes must be sent back for warranty processing
within two months (starting from the day of delivery).
3.4
Storing the dAg ring electrode
The electrode head must be stored as follows to protect it from water, sol-
vents, dust and mechanical influences:
1
Screw the protective cap
onto the electrode head
2
Store the electrode in the storage vessel. When doing so, ensure that
the diaphragm
is immersed in the corresponding storage solu-
tion.
NOTICE
We recommend using the reference electrolyte as a storage solu-
tion.
3
Close the filler opening