Maintenance
■■■■■■■■
9
5 Maintenance
5.1
Metal electrode – Changing/refilling the electrolyte
1
Open the filler opening.
2
Use a plastic pipette to empty the electrode.
3
Rinse the inside of the electrode with the new electrolyte.
4
Fill the electrode with electrolyte up to the filler opening.
5
Close the filler opening if the electrode is not used immediately.
6
Immerse the electrode in electrolyte solution overnight.
Then, the electrode is ready for use again.
5.2
Assessing a metal electrode
Carrying out a standard titration
1
Fill a 100-mL beaker with 50 mL of deionized water.
2
Add 2 mL of hydrochloric acid (
c
=0.1 mol/L).
3
While stirring this mixture, titrate it with silver nitrate (
c
=0.1 mol/L)
under the following conditions:
Method
DET U
Dosing rate
Max.
Signal drift
50 mV/min
Min. waiting time
0 s
Max. waiting time
26 s
Measuring point dis-
tance
4
Min. increment
10.0 µL