Preparing the dAg ring electrode
■■■■■■■■■■■■■■■■■■■■■■
6
■■■■■■■■
4 Installation
4.1
Preparing the dAg ring electrode
1 Filling with reference electrolyte
Open the closure of the filler opening
and, if necessary, fill ref-
erence electrolyte up to the filler opening.
2 Cleaning the electrode
■
Rinse the electrode with distilled water.
■
If there is excess contamination on the metal ring, clean it with a
moist towel and toothpaste or with the polishing set
(6.2802.000).
■
If necessary, use a suitable solvent to degrease the electrode.
NOTICE
The electrode should be rinsed before each measurement.
Frequent abrasive cleaning is not recommended.
3 Cleaning a diaphragm that is clogged with silver chloride
■
■
Immerse the electrode in a concentrated ammonia solution or 7%
thiourea solution in 0.1 mol/L of HCl for approximately 2 hours.
■
Rinse the electrode with distilled water.
■
Refill with 1 mol/L of KNO
3
as a reference electrolyte.
■
Immerse the electrode in 1 mol/L of KNO
3
for several hours.
4 Connecting the electrode
■
Unscrew the protective cap
.
■
Position the cable connection on the electrode head such that the
slot in the cable connection is on the guide lug of the electrode
head.
■
Push the socket in the cable connection into the plug inside the
electrode head.