Installation
■■■■■■■■
7
■
Push the outer ring of the cable connection over the electrode
head.
Ensure that the guide lugs in the electrode head are in the
grooves of the cable connection.
■
Push the cable connection onto the electrode head until it stops
and rotate the outer ring until it snaps in place.
NOTICE
To remove the cable, first release the outer ring and then carefully
pull the cable connection from the electrode head.
When doing so, be sure not to pull on the cable itself but the cable
connector instead.
4.2
Mounting the electrode
The electrode must sit securely in the titration head.
NOTICE
For automatic procedures, ensure that the cables have enough room
to move.
During the titration, it is important that the solution is mixed well. The stir-
ring rate should be high enough to form a small vortex. If the stirring rate