6 Troubleshooting
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
49
6 Troubleshooting
6.1
Problems and their solutions
Problem
Cause
Remedy
The baseline has a
large amount of
noise.
MCS – The
CO
2
adsorber cartridge CW
is depleted.
Replace the CO
2
adsorber cartridge CW
.
MCS – The vacuum pump is
defective.
Contact Metrohm Service.
The pressure in the
system markedly
increases.
The MSM is blocked.
■
Regenerate the MSM
.
Note: 6.2821.180 pump tubing connection
with filter must be used.
Peak areas are lower
than expected.
MCS – The CO
2
suppressor
is not connected.
Connect the CO
2
suppressor.
The peristaltic pump
is pumping too little.
Peristaltic pump – Contact
pressure too weak.
Correctly set the contact pressure
the contact pressure correctly", page 20)
.
Peristaltic pump – Filter
blocked.
.
Peristaltic pump – Pump
tubing defective.
Replace the pump tubing
.
MSM – Insufficient
flow of regeneration
solution or rinsing
solution.
Peristaltic pump – Contact
pressure too weak.
Correctly set the contact pressure .
Peristaltic pump – Filter
blocked.
.
Peristaltic pump – Pump
tubing defective.
Replace the pump tubing.
The background
conductivity is too
high.
The MSM is not connected.
Connect the MSM
MSM – No or insufficient
flow of regeneration solu-
tion or rinsing solution.
Check the flow of the regeneration solution
and of the rinsing solution .
MCS – The CO
2
suppressor
is not connected.
Connect the CO
2
suppressor.