3 Installation
942 Extension Module Vario SeS/PP (2.942.0500)
■■■■■■■■
9
3 Installation
3.1
Setting up the instrument
3.1.1
Packaging
The instrument is supplied in protective packaging together with the sepa-
rately packed accessories. Keep this packaging, as only this ensures safe
transportation of the instrument.
3.1.2
Checks
Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.
3.1.3
Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for
operation and free of vibrations and which provides protection against
corrosive atmosphere and contamination by chemicals.
The instrument should be protected against excessive temperature fluctua-
tions and direct sunlight.
3.2
Transport locking screws
To avoid damage to the vacuum pump drive during transport, the pump is
secured with transport locking screws. These are located at the rear of the
instrument and labeled with
Transport security screws
.
Remove these transport locking screws before the initial start-up.
Accessories
For this step you need:
■
4 mm hex key (6.2621.030)
Removing the transport locking screws
1
Remove all of the transport locking screws with the hex key.
Store the transport locking screws in a safe place. Reinsert the transport
locking screws each time you transport the instrument a significant dis-
tance.