3.5 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
942 Extension Module Vario SeS/PP (2.942.0500)
3.5.2.3
Connecting the regeneration solution
The regeneration solution is connected to the
regenerant
capillary.
Connecting the regeneration solution to the peristaltic pump
For this step, you need the following accessories:
■
Pump tubing (6.1826.420)
■
Tubing olive with filter and locking nut (6.2744.180)
■
Tubing olive (6.2744.034)
■
Tubing cartridge of the peristaltic pump
1
Prepare a tubing cartridge of the peristaltic pump for the regenera-
tion solution
2
Connect the capillary labeled
regenerant
to the outlet of the pump
tubing using a pressure screw (6.2744.070).
3
Connect the PTFE capillary from the regeneration solution bottle to
the inlet of the pump tubing.