6 Troubleshooting
889 IC Sample Center
■■■■■■■■
43
6.1.6 Electronics
Problem
Cause
Remedy
Error 280, 282, 283,
284
An electronic error has occur-
red in the EEPROM.
1. Switch off sample changer.
2. Wait a few seconds.
3. Switch on sample changer.
If the error occurs again, please call the
Metrohm Service.
Error 290
When initializing the sample
changer at least one critical
malfunction has been found.
No injections can be carried
out.
1. Exit MagIC Net.
2. Switch off sample changer.
3. Ensure that the cable between the
sample changer and the PC is connec-
ted correctly.
4. Switch on sample changer.
5. Start MagIC Net.
If the error occurs again, please call the
Metrohm Service.
6.2
Analytical problems
Analytical problems such as poor reproducibility and carry-over can occur in
any chromatography system. The cause is usually difficult to find. First of all,
you should attempt to determine whether the problem is caused by the
Autosampler or another component of the system.
Please take into consideration the fact that analytical problems can also be
caused by external influences such as temperature or light-sensitive sample.
Make sure that the application was able to be performed previously without
problems and that no changes have been made to the system.
Some causes and possible solutions for analytical problems are listed below.
Contact your service point for additional help.
6.2.1 Autosampler
Problem
Cause
Remedy
The reproducibility
does not correspond to
the specifications.
There is air in the flow path.
Carry out a standard washing procedure
(in MagIC Net
™
under
Manual con-
trol
).