5 Handling and maintenance
889 IC Sample Center
■■■■■■■■
35
Note
If you are using the sample racks with 12 or 48 vials, then you must
adjust the setting of the needle height to > 2 mm. This will prevent
the needle from touching the bottom of the vial.
18
Perform a rinsing procedure in order to clean the new needle. Start the
Wash
function. To interrupt the rinsing procedure, click on
Stop
.
19
Note
The needle settings may need to be adjusted.
In MagIC Net
™
, open the configuration settings of the
889 IC Sample
Center
and select the
Needle
tab.
Perform the necessary setting.
5.3.3
Replacing the air needle
Replace the air needle as follows:
1
Replace the sample needle,
.
2
Undo the fastening nut (made of chrome) for the air needle.
3
Undo the fastening nut (made of chrome) for the adjusting screw.
4
Remove the air needle.
5
Tighten the height adjustment nut on the fixing nut (made of chrome).
The thread of the height adjustment nut must match the lower part of
the fixing nut. The O-ring must be properly seated in the fixing nut.
6
Insert the air needle.
7
Mount the sample needle.
8
Select the correct needle height for the new needle in the manual
operation feature of MagIC Net
™
on the
Needle
tab.