4.3 Full loop injection
■■■■■■■■■■■■■■■■■■■■■■
22
■■■■■■■■
889 IC Sample Center
The syringe aspirates a rinsing volume of sample solution out of the
sample vial and fills the sample line with sample. The rinsing liquid is
expelled.
3 Switching the injection valve to LOAD
The injection valve is switched over to the
LOAD
position. This causes
an homogenous sample plug to be present at the inlet to the sample
loop.
4 Filling the sample loop
The sample loop is quantitatively filled by a multiple of the loop volume
being pumped through the loop.
■
3 × loop volume with sample loops
≤
100 µL