5 Handling and maintenance
889 IC Sample Center
■■■■■■■■
33
5.3.2
Replacing the sample needle
Figure 14
Replacing the sample needle
Replace the sample needle as follows:
1
Open the
Manual control
in MagIC Net
™
.
2
Under device selection,
889 IC Sample Center
, select the tab
Nee-
dle
. Under Position/Input, select
Exchange
position and click on
[Start]
.
Now a message window is displayed.
3
Remove the sample rack and click on
[OK]
.