Technical specifications
XI - 62
Virtual digital oscilloscopes, 60 MHz
Technical specifications
Vertical offset
Only the assigned tolerance or limit values are guaranteed (after
30 minutes to adapt to temperature).
Values without tolerances are given for information purposes only.
Characteristics
Specifications
Observations
Number of
channels
2 channels: CH1 - CH2
BNC entries
Types of
entry
Class 1, common masses
Bandwidth
to - 3 dB
≥
60 MHz on all vertical ranges
from 5 mV to 5 V/div.
≥
20 MHz on ranges 10 V/div. to
100 V/div. ( the bandwidth limitation
for these calibres is related to the
HF power limitation in the entry
capacity attenuation network)
Measured on a charge of 50
Ω
with a 6 div. amplitude signal
Dynamics of vertical
offset
±
10 div. on all the ranges
Entry
Coupling
AC, DC, GND
Bandwidth
Limit
at 15 MHz, 1.5 MHz and 5 kHz
one bandwidth limit per channel
Rise
time
≤
5 ns (60 MHz) on vertical ranges
from 5 mV to 5 V/div.
depending on the model
Cross-talk
between channels
DC at 50 MHz
≥
40 dB
same sensitivity on both channels
Tolerance
ESD
± 2 kV
Response
to rectangular 1 kHz
and 1 MHz signals
Overshoot
< 3 %
Aberrations
< 3 %
Positive or negative overshoot
Accuracy
of vertical
ranges
Vertical
resolution
Accuracy of vertical
measurements
Accuracy of vertical
offset
±
2 % (on a 6 div. amplitude signal)
±
0.4 % of the full scale
±
[2 % (reading - offset)
+ Precision of vertical offset
+ (0.05 div.) x (volt/div.)]
±
[0.01 x (offset value)
+ 4 mV + (0.1 div.) x (V/div.)]
Sequence of vertical ranges
1 - 2 - 5
Variation in bounds (no continuous
variable rate)
Probes
Attenuation coefficients of the probe
and its graphical representation are
taken into account in the display
Vertical
ZOOM
function on an
acquire or backed up curve
no vertical Zoom
maximum entry Voltage
400 Vpk (DC + peak AC at 1 kHz)
without the sensors
Electrical
safety
300 V, CAT II without sensors
Entry
Impedance
1 Mohm
±
1 % 13 pF ± 2 pF
Display
modes
"Multi-window" with the possibility of
displaying the f(t) trace, the FFT
and the XY mode simultaneously
Default display: control
trace window
Double time base type display,
even in real time