MI 3394 CE MultiTesterXA
Single
tests
29
Ipe Leakage measurement procedure
Select
the
Ipe Leakage
function.
Set test parameters / limits.
Connect device under test to mains test socket.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.33: Examples of Ipe Leakage measurement results
4.1.10 Touch
Leakage
Figure 4.34: Touch Leakage test menu
Test results / sub-results
Itou ............ Touch Leakage current
P ............... Power
Test parameters
Duration
Duration [Off, 2 s ... 180 s]
Change status
Change [YES, NO]
YES: The instrument measures leakage current in two sequential steps
with 5 s delay in between. The phase voltage is firstly applied to
the right live output of the mains test socket and secondly to the
left live output of the mains test socket.
NO: The phase voltage is applied only to the right live output of the
mains test socket.
Output connections [Socket L,N – PE,P/S]