In pre-test an external voltage between C1/P1 and
C2/P2 terminals was detected.
The measurement was cancelled. Press OK to continue.
In pre-test a too high external voltage was detected
between P and PE terminals. The measurement was
cancelled. Press OK to continue.
In pre-test a too high external voltage was detected
between ISO/SUB and PE terminals. The measurement
was cancelled. Press OK to continue.
In pre-test a possible high leakage current was
detected. It is likely that a dangerous leakage current
(higher than 3.5 mA) will flow after applying power to the
device under test.
Select
YES
to proceed with or
NO
to cancel
measurement.
The measured leakage (Idiff, Ipe, Itouch) current was
higher than 20 mA. Measurement was aborted. Press
OK to continue.
The load current exceeded the highest upper limit of 10
A for the Discharging time test. Press OK to continue.
The instrument is overheated. The measurement can’t
be carried out until the icon disappears. Press OK to
continue.
The device under test should be switched on (to ensure
that the complete circuit is tested).
Test voltage in Insulation resistance measurement is too
low.