MI 3394 CE MultiTesterXA
Single
tests
16
Test results / sub-results
U ............... measured test voltage
I ................. test current
Test parameters
DC test voltage
U test [500 V ... 6000 V in steps of 50 V]
Duration
t end [Off, 1 s ... 120 s]
Test limits
High limit (I)
H limit [0.05 mA ... 10.0 mA ]
Low limit (I) L limit [Off, 0.05 mA ... 10.0 mA]
Test circuit
Figure 4.9: HV DC measurement
HV DC measurement procedure
Select the
HV DC
function.
Set test parameters / limits.
Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).