![Metal Samples Company MS3500E Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/metal-samples-company/ms3500e/ms3500e_manual_1776981047.webp)
43
It is important to note that the ER Meter Prover is only a check probe used to verify general instrument
operation. It is not a calibration standard. The measured value may differ from the listed value as
explained in step 3 above.
It is also important to note that each ER Meter Prover is serialized to match the instrument it is shipped
with, and its measured value may differ if it is used with another meter. If it is necessary to use an ER
Meter Prover with another meter, it is recommended that you establish a baseline value for the ER
Meter Prover with the other meter.
Содержание MS3500E
Страница 2: ... ...
Страница 40: ...36 Figure 8 Element Selection Guide Element Life ...
Страница 41: ...37 Figure 9 Element selection guide response time ...