41
Appendix A - MS3500E Specifications
Model
MS3500E - Remote ER Data Logger (Ordering # IN3500)
MS3510E - Remote ER Data Logger w/ 4-20mA Current Loop Output (Ordering # IN3510)
Physical Data
Instrument Weight:
11.94 lb. (5.42 Kg)
Total Weight w/ Accessories:
13.64 lb. (6.19 Kg)
Instrument Dimensions:
11.50"H x 8.94"W x 4.00"D (29.21cm x 22.71cm x 10.16cm)
Case Specifications:
NEMA 4X (IP 66)
Mounting Specifications:
10.75"H x 6"W (27.31cm x 15.24cm) Bolt Pattern
0.3" (0.76cm) Diameter Bolt Holes
Operating Temperature:
32° to 122°F (0° to 50°C)
Storage Temperature:
-4° to 158°F (-20° to 70°C)
Performance Data
Measurement Type:
ER measurement using any standard ER probe type (Wire Loop,
Tube Loop, Cylindrical, Flush, Strip, etc.)
Range:
0-1000 digits representing 0-100% of probe life
Resolution:
1 digit
Cycle Time:
1 Hour to 99 Days
Electrical Data
Power Requirements:
Six 1.5V AA Batteries
Max. Probe Cable Distance:
10 ft (3.05 m)
Output Specifications:
RS-232 Output in Comma-Delimited ASCII Text Format
4-20mA Current Loop Output (MS3510E Only)
Class I, Division 1Groups A, B, C, and D
Temperature Code T3C
Class I, Zone 0,
Group IIC, T3C
Conforms to ANSI/UL Std. 913
Special Features
• Microprocessor-based electronics
• Data storage capacity of 3100 readings, with battery backup
• Menu-driven interface using a 2-key keypad and a 2-line LCD display
• Low-battery detection
Accessory Items
10' Probe Cable, Meter Prover, Communications Cable and Connector, Current Loop Connector
(MS3510E only), Operation Manual
Содержание MS3500E
Страница 2: ... ...
Страница 40: ...36 Figure 8 Element Selection Guide Element Life ...
Страница 41: ...37 Figure 9 Element selection guide response time ...