AEROCET 531S Manual
Page 23
AEROCET 531S-9800 Rev A.docx
7.3.
Comma Separated Value (CSV)
The CSV report will be generated when the
REPORT
setting is set to
CSV
. The CSV format for
MASS
and
COUNTS
is
determined by the
MEASURE
setting. Both formats are fixed field length.
7.3.1.
COUNTS
Format
Mass Data Report
2013-01-17 16:56:26
Serial Number, N12345
Time,0.5(M3),1.0(M3),5.0(M3),10 (M3),AT(C),RH(%),Location,Seconds,Status
2013-01-17 15:49:06,03560226,00399152,00013069,00020,027,001,060,000
CSV Fields
Field
Parameter
Example Value
Notes
1
Date and Time
2013-01-17 15:49:06
2
Channel 1 Size 0.5 (TC, /L, CF, M3)
03560226
Note 2
3
Channel 2 Size 1.0 (TC, /L, CF, M3)
00399152
Note 2
4
Channel 3 Size 5.0 (TC, /L, CF, M3)
00013069
Note 2
5
Channel 4 Size 10 (TC, /L, CF, M3)
00006358
Note 2
6
AT (C, F)
+020
Note 2 & 3
7
RH (%)
027
Note 3
8
Location
001
9
Seconds
060
Note 4
10
Status
000
Note 5
Status Bits
Bit
Value
Condition
0
OK (no alarms or errors)
0
1
Not used
1
2
Not used
2
4
Not used
3
8
Not used
4
16
Low battery
5
32
Sensor error
Notes (for table above)
1. CSV header included for multiple record transfers like All Data (2)
or New Data (3).
2. Units determined by product setting.
3. Temperature and RH will be null (,,) if Temp/RH probe is not
attached
4. The sample time is fixed at 60 seconds.
5. Status bits combinations are possible. See the Status Bits table.
For example, 48 (00110000B) = Low battery and Sensor error.