Docking Station Set Up Manual 83529-9800 Rev A
Page 11
8. Connect the computer or smartphone to the same network that you selected for
the Docking Station WiFi.
9. In the web browser you should now be on the new IP address that the network
assigned to the dock.
10. Navigate to the uC Console tab and set the baud rate from the drop down tab.
This must match the Dock Baud of the measurement device. Default Dock Baud
for all devices is 38400.
Figure 10 - Microcontroller Console
11. With a Met One Instruments, Inc. handheld device resting on the dock and the
network selector switch set to
WiFi
you can now communicate with the device
using serial commands. A TCP/IP connection is made using the assigned IP
through Port 23.
4.3.
Ethernet Port Setup and Configuration
The Docking Station Ethernet port must be configured with some drivers by the user:
4.3.1. Setting the Static IP Address of the Docking Station
1. You will need to obtain a Static IP Address and Network Mask from your network
administrator.
2. Move the WiFi/Ethernet switch on the rear of the Docking Station to Ethernet.
3. Connect CAT5 Ethernet cable between local network and Ethernet connector on
the back of the Docking Station.