ID
Area
Reconditioning instructions
5 Base
5.1
Welds
Check all welds carefully for signs of cracks, corrosion
and movement. Ensure good lighting and preferably use a
magnifying glass, as small cracks can be difficult to detect
on black lacquered bases. Make sure the base is stable.
Fill in any varnish damage with touch-up paint to avoid
corrosion.
5.2
Bolted joints
Check and tighten all screw joints. Replace screws that have
damaged key handles or threads.
6 Brakes
6.1
Function
Replace the brake feet. Check that the brake is working
properly.
6.2
Plastic plugs
Check that guide bushings and plastic plugs are properly
secured. Replace worn or damaged parts.
6.3
Pinch protection
Check that any pinch protection is securely mounted.
If there is no pinch protection on the brake legs, it is
recommended that this be installed before the device is put
back into service.
7 Castors
7.1
Function
Check castors and jumpers for wear or play. Wear or play
may indicate defective bearings. Check that the castors
rotate well and the housings rotate. Replace worn or
damaged parts.
7.2
Attachment
Check that the castors’ screws are intact and tightened.
Replace screws that have damaged key handles or threads.
8 Seat frame
8.1
Mechanics
Carefully check all welded parts for cracks and paint
damage. In particular, check around the mounting of
actuators for the seat tilt and at the holes for mounting leg
supports, as these areas are particularly subject to load.
Ensure good lighting and preferably use a magnifying glass,
as small cracks can be difficult to detect on black lacquered
structures. Fill in any varnish damage with touch-up paint to
avoid corrosion.
8.2
Bolted joints
Check and tighten all screw joints. Replace screws that have
damaged key handles or threads.
8.3
Plastic plugs
Check that guide bushings and plastic plugs are properly
secured. Replace worn or damaged parts.
8.4
Controls
Make sure that all knobs and handles on the seat frame are
working and that they lock correctly.
53
Instructions for use for medical work chairs in the REAL 9000 PLUS device range
Servicing & reconditioning
21-08742-EN-01