8. Servicing & reconditioning
This section describes the inspection and troubleshooting of the device, information about
battery replacement and instructions for reconditioning the device. The troubleshooting guide
is aimed at all users of the device, and the section is otherwise aimed at those who handle
these parts of the device lifecycle professionally.
PLEASE NOTE! The user must not sit in the device while it is being reconditioned or serviced or
during maintenance.
Inspection during service
A thorough visual inspection must be
performed of the device’s main components
in order to guarantee the patient’s safety.
The main components of the device include
base, brake, lifting mechanism, seat frame,
back recliner and armrest mechanism. The
inspection must include welds, lockable
functions and settings.
When checking the PLUS base, always
inspect the weld between the base’s front
and rear legs. Any indication of cracks in
the material or paint in this area should be
regarded as reason to replace the base,
or the device. Ensure good lighting and
preferably use a magnifying glass, as small
cracks can be difficult to detect on black
lacquered bases.
In case of reduced braking action, check both
the stability of the base and the condition of
the brake rubber. If the device is not resting
on all four wheels in the unbraked position,
the device must be withdrawn from service
immediately. Check the balance of the base
in several locations and in several directions
to exclude any influence from the floor.
Warning! Repairs and other technical
measures may only be carried out
by personnel authorised by Mercado
Medic. If this is not followed, the
CE marking no longer applies and
Mercado Medic may no longer have
full device liability.
Warning! As a user, you should
contact your prescriber, technical
aids centre or distributor if the
device shows reduced or altered
performance. A device showing a loss
of or change in performance must
immediately be taken out of use in
order to avoid an accident. The device
must not be used again until an
authorised technician has examined
the device.
45
Instructions for use for medical work chairs in the REAL 9000 PLUS device range
Servicing & reconditioning
21-08742-EN-01