ID
Area
Reconditioning instructions
3 Electric lifting mechanism
3.1
Noise
Listen for noise in the lifting mechanism. Replace the lifting
mechanism in case of noise. Load the device and check that
the lifting mechanism does not slip when starting from the
bottom position. Defective lifting mechanisms can be sent
to Mercado Medic for repair.
3.2
Rotation lock
Make sure that the lifting mechanism does not have any
play when rotating or loading from side to side. Replace
the lifting mechanism in case of play. Defective lifting
mechanisms can be sent to Mercado Medic for repair.
3.3
Lifting columns
Raise the lifting mechanism to the highest position. Wipe
the lifting column of dust and dirt with a dry cloth. Then
lubricate the lifting column using a thin layer of Teflon or
silicone-based grease.
3.4
Attachment
Check that the seat is facing straight ahead relative to the
base. Check that the lifting mechanism attachments to the
seat frame are correctly assembled and tightened. Replace
screws that have damaged key handles or threads.
3.5
Cabling
Inspect cables for any signs of wear, pinching or crushing.
4 Gas spring-operated lifting mechanism
4.1
Rotation lock
Make sure that the lifting mechanism does not have any
play when rotating or loading from side to side. Replace the
lifting mechanism in case of play.
4.2
Lifting columns
Raise the lifting mechanism to the highest position. Wipe
the lifting column of dust and dirt with a dry cloth. Then
lubricate the lifting column using a thin layer of Teflon or
silicone-based grease.
4.3
Attachment
Check that the seat is facing straight ahead relative to the
base. Check that the lifting mechanism attachments to the
seat frame are correctly assembled and tightened. Replace
screws that have damaged key handles or threads.
4.4
Lock ring
Check that the lock ring under the bottom is securely
fastened.
4.5
Gas spring lever
Check that the gas spring locks properly and that the height
does not slowly change under load. If the height slowly
changes under load, adjust the wire nipple. Check and
replace damaged wires and sheaths.
52
Instructions for use for medical work chairs in the REAL 9000 PLUS device range
Servicing & reconditioning
21-08742-EN-01