Reconditioning instructions
Do not use high pressure washers when cleaning the device. For instructions on everyday
cleaning, see 3. Caring for the device.
The following points should be performed during reconditioning to ensure patient safety:
ID
Area
Reconditioning instructions
1 Cleaning
1.1
Mechanisms
Clean mechanical parts (not the actuator) using a clean,
damp cloth with mild detergent (pH 7-12), surface
disinfectant or use a steam cleaner (max. 8 bar).
PLEASE NOTE! Do not wash the device with water or other
liquids and chemicals.
1.2
Electronics and cables
Remove dust with a dry cloth.
1.3
Operating controls
Wipe the control controls with a cloth lightly dampened
with disinfectant. This is so as not to pass on any possible
infection.
1.4
Actuator and lifting
mechanism
Moving parts are lubricated and should therefore not be
cleaned. For lifting mechanisms, see special instructions
under sections 3 and 4.
1.5
Padded parts
Remove and discard all padded parts. Wash any covers
according to the washing instructions.
1.6
Castors
Clear the castors of hair and dust.
Long-term storage of batteries
During longer periods of storage, the battery
is automatically disconnected after seven
days. The battery may also be forced into
unplugged mode by plugging a 9V power
supply (part number 804823) into the
charging socket and waiting for at least 5
seconds before disconnecting the power
supply. If the handheld controller is used,
all of the indicator lights will flash when
the battery is disconnected and the power
supply can then be disconnected.
When the battery is disconnected, it must
be charged every four months in order to
retain its functionality, see 2.3. Charging.
If these charging instructions are followed,
the battery will maintain sufficient capacity
for 5 years.
To reconnect the battery to the system
again, connect the battery charger to the
charging outlet and wall socket. Wait at
least five seconds before disconnecting
the battery charger. The battery is now
activated again and the system is ready for
use. We recommend leaving the charger in
place and fully charging the battery after the
device has been stored with the battery in a
disconnected state.
50
Instructions for use for medical work chairs in the REAL 9000 PLUS device range
Servicing & reconditioning
21-08742-EN-01