Page 3 of 8
Data Sheet CPC3000
Specifications
CPC 3000
Pressure ranges
Gauge: 0 ... 5 up to 0 ... 1500 psig (0 ... 0.35 to 0 ... 103 bar)
Absolute: 0 ... 14.5 to 0 ... 1515 psia (0 ... 1 to 0 ... 104 bar)
Bi-directional: 5 ≤ span ≤ 1515
Pressure types
Absolute, gauge or bi-directional ranges
Uncertainty
0.025% full span for all standard pressure ranges specified above. Uncertainty (k=2) includes
hysteresis, linearity, repeatability, reference standard, drift and temperature effects over the calibrated
range for the calibration interval specified, with periodic re-zeroing.
Optional Intelliscale Uncertainty
Gauge
Transducer Range (psig)
Total Uncertainty
(1)
(cal interval)
0 ... 14.5 to 0 ... 1500
0.025% IS-50
(1)
(365 days)
Absolute
Transducer Range (psia)
Total Uncertainty
(1)
(cal interval)
0 ... 14.5 to 0 ... 1515
0.025% IS-50
(1)
(365 days)
Bi-directional
Transducer Range (psi)
Total Uncertainty
(1)
(cal interval)
-15 ... 145 to -15 ... 1500
0.025% IS-50
(1)
(365 days)
5 ≤ span < 145
0.025% of Full Span (365 days)
(1) 0.025% IntelliScale-50 (0.025% IS-50): Uncertainty from Min to 50% of Max = (0.025% x 50%
x Max) or 0.025% of Reading from 50% to 100% of Max.
Optional barometer uncertainty
0.02% R from 8.5 to 17 psia for 365 days
Compensated temperature range
15 to 45
°C
Calibration interval
365
days
Pressure units: English
psi, psf, oz/si, tons/sq in, tons/sq ft, atm, inHg @0C, inHg @60F, mTorr, Torr, inSW, ftSW, inH2O @4C,
inH2O @20C, inH2O @60F, ftH2O @4C, ftH2O @20C, ftH2O @60F
Pressure units: Metric
mbar, bar, g/sq cm,
kg/sq cm, dyn/sq cm, pascal, hPa, kPa, MPa,
mmHg @0C, cmHg @0C, mHg 0C, mSW, mmH2O @4C, cmH2O @4C, mH2O @4C, mmH2O @20C,
cmH2O @20C, mH2O @20C
Pressure units user defined
2 (multiplier from psi or Pascal)
Slew rate
3 seconds to stable flag (+/-0.025% full scale pressure) for a 10% pressure change typical into
150cc volume at pressures greater than 5 PSI. Larger volumes can lengthen this time. Controlling to
pressures less than atmosphere can lengthen this time.
Overshoot
<1
% FS,
in high speed mode
Control stability
0.003% of range
Pressure ports
7/16-20 Female SAE threaded ports for Measure/Control, Exhaust, Reference, and Supply.
Barometric Reference port is a hose barb.
Filter elements
40 micron filter element included in each pressure port (excluding the optional barometer and the
reference port on a gauge sensor)
Permissible pressure media
Clean, dry, non-corrosive gases
Parts exposed to pressure media
6000 series aluminum, 316 SS, brass, Teflon, Urethane, Silicone, RTV, Silicone grease, PVC, Epoxy, Ceramics
Overpressure protection
Pressure relief valves
Instrument mounting
Desk top with bezel and handle, or optional rack mount kit.
Display
7.0" color LCD
Resolution
Six significant digits
Warm-up time
approx. 15
min
Digital interface
IEEE-488, Ethernet, USB
Power supply
100 - 240 VAC, 50/60 Hz, 700 mA max
Maximum pressure, supply port
110 … 120 %FS
Pneumatic overpressure protection
internal relief valves
Operating temperature
0 … 50
°C
Storage temperature
0 ... 70
°C
Air humidity
0 … 95 (% relative humidity without moisture condensation)
Operating position
Negligible, can be removed with re-zeroing
Weight
<20 lbs.(<9.1 kg) with all internal options
Dimensions
5¼ h x 8⅜ w x 12 d in (133 x 213 x 305 mm)
CE-mark
Conformity certificate
Calibration
NIST traceable calibration certificate included, A2LA certification