28
06/2021
Shaking Device
5.5
Faults, Warning & Error Messages
CAUTION
If the shaking device is used strangely, for example, uncontrolled shaking, you may
suffer bruising injuries to your hands. Disconnect the appliance from the power supply
by pulling out the mains plug.
Description
Cause
Action
Motor lock-up error
■
Restart movement,
■
Reduce load.
E-111
130
rpm
Software error
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-132
130
rpm
Memory access error
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-135
130
rpm
CRC error (internal memory error)
■
Restart appliance
■
Contact Service
in case of frequent occurrence
E-139
130
rpm